UOL Spray Dryer SOP

By

Standard Operating Procedure

Introduction:

The spray dryer (Figure 1) is located in the Unit Operations Laboratory (MEB 3969). Spray dryers are used to form dry powder from a liquid or slurry by rapidly drying with hot gas. This spray dryer uses a spray nozzle to disperse the liquid or slurry into a spray with a controlled drop size.

 

Before operating the Bowen Spray Dryer, familiarize yourself with the plumbing, display gauges, control panel, pump, pump drive, and blower. The Bowen Engineering Spray Dryer is a pilot-scale device commonly used for drying thermally sensitive materials. In the Senior Lab (MEB 3969), it has been used for drying a variety of solutions including soap, salt, and beer. Before using this Spray Dryer, verify that any solution pumped through the system will not clog the lines or the spray nozzle. Check with the lab manager before preparing a new solution.

Figure 1. Spray Dryer.

 

Health & Safety Warnings:

  • Safety glasses must be worn when operating this equipment.
  • The manual gas control valve must be closed after each run.
  • Do not remove the drive shaft guard.
  • To prevent overheating, the gas burner will not light until the blower is powered
    on.
  • Use caution when opening the spray dryer door during operation as liquid may
    escape the drying chamber.
  • Caution – exterior surfaces may be hot!
  • Rotating equipment hazard: Use caution when in close proximity to the pump.

 

Procedures:

Pre-Run Steps

  1. Mix feed solution in tank(s) located behind the spray dryer (refer to Figure 2). Verify that the feed tank valves remain closed until you are ready to run.
  2. Set (open) valves on feed tanks to feedstock solution to pump as desired (refer to Figure 2).
  3. Using a wrench, remove the pump drain plug until feedstock/water flows out of the pump housing. Then, replace the plug, tighten it finger tight, and then tighten it ¼ turn with a wrench (Figure 3 shows the pump).
  4. If feedstock does not flow out of the pump valve, contact the lab manager.
  5. Using a small crescent wrench, open the main gas pipe valve ½ turn so that the valve headpiece is in line with the pipe. The valve is located on the rear of the spray dryer (refer to Figure 4).
  6. Open the main air supply valve located on the right side of the dryer ½ – 1 turn.
  7. Set the air pressure regulation to 20 psi. The air pressure regulator is located on the lower front panel. You may adjust the air pressure as required as you fine-tune your operating parameters. The air flow rate may be adjusted by adjusting the air pressure regulator.
  8. Turn on the main power switch located on the upper left of the control panel.

Check the Chamber

  1. Open the chamber door. The chamber’s light switch is located on the right side of the control panel. Verify that there are no obstructions in the chamber. If the chamber requires cleaning, use the hose located behind the spray dryer to clean the system. Cleaning supplies are in the stock room. It is your responsibility to clean the spray dryer after each use.
  2. Verify that the spray nozzle (Figure 5) is connected to the air and feed lines.
  3. Close the chamber door.
  4. Latch both chamber door fasteners to secure the door.

Figure 2. Mixing tank. In the left background is a larger tank used for very extended runs. When ready to run adjust the valves seen leaving the tank to valve off the other tank and to connect to the spray dryer.

Figure 3. View of the pump.

Set pump speed

  1. To turn on the pump, press the red button labeled “Pump Switch” on the Opto interface.
  2. Once the pump has started you can adjust the power of the pump by changing the Pump Power percentage through the Opto interface.
    • The flow rate of the pump should be displayed just below the pump power.

Figure 4. The valve for the gas line is seen immediately in front of the gray-colored gas meter.

Figure 5. Spray nozzle and chamber light (left-hand panel). The light on-off switch is shown in the middle panel. See also the burner on/off switch. the right-hand panel shows the main power on-off switch

Figure 6. The pump is controlled by Opto 22 equipment. You will control the pump from the GUI.

You can confirm the spray is adequate before closing he chamber door (Refer to Figure 7).

Figure 7. With the pump on (at left), you can verify that there is a stream of fluid. With the pump and the air on (at right), you can confirm that the spray is appropriate and that there are no obvious issues with the nozzle.

Stopping the pump

To stop the pump, push the now green “Pump Switch” button again on the Opto interface.

Starting the blower

To start the blower, press the red button labeled “Blower Switch” on the Opto interface.

Lighting the burner

  • Note – the blower must be running in order to ignite the burner (Figure 8).
    1. Verify that the main gas valve is open.
    2. Push the black start button (burner control – see Figure 4) located directly above the fan start/stop control. Verify that the burner is lit via the observation port (Figure 9) on the burner assembly.

Figure 8. You will control the blower from the GUI.

Sample Collection

  1. Sample material is collected in the bell jar located at the bottom right of the spray dryer. The jars will thread in below the cyclone separator (Figure 10).
  2. Additional bell jars and spray nozzles are in the cabinet adjacent to the spray dryer.

Figure 9. Burner observation port.

Figure 10. Collection jars thread in below the tapered cyclonic separator. At the left, the the main compressed air valve is shown.

 

Shutdown:

  1. Set the pump speed to Zero. Stop the pump.
  2. Turn off the burner.
  3. Turn off the blower.
  4. Turn off the main power switch.
  5. Close the main gas valve.
  6. Close the feed valves (ball valves) that control the feed flow from the feed tanks to the pump.
  7. Drain and rinse the feed tanks as needed.
  8. Determine the moisture content of the sample that you have acquired. This will mean you will need to have tared the sample jar.
  9. Rinse the spray dryer chamber with tap water.
  10. Clean the work area.
  11. If you plan on leaving feed solution in the tanks overnight, it must be labeled.

Associated SOP Files